Ophthalmic Eye Sphere and Conformer Sterilization Instructions |
|
|
| |
| Gulden Ophthalmics Eye Spheres and Conformers are non sterile. Eye Spheres and Conformers may be sterilized with ethylene oxide (EtO) under the following conditions:
|
Temperature
Pre-Vacuum
Relative Humidity
Humidty Dwell
Sterilant Gas
EtO Concentrate
Exposure time
Post Vacuum (2 times)
Mechanical Aeration temperature
|
130 ± 5°F / 54 ± 3°C
24 - 26 inHg / 610 - 660 mmHg
60 ± 10% RH
30 - 45 minutes
100 % EtO
600 ± 30 mg/Liter
2 hours ± 5 minute
24 - 25 inHg / 610 - 660 mmHg
110 - 124°F / 43 - 51°C
|
|
Residuals after 48 hours minimal Aeration time
|
EO < 250ppm
ECH < 250ppm
EGly < 5000ppm
|
|
Residuals after 72 hours recommended Aeration time
|
EO < 250ppm
ECH < 250ppm
EGly < 5000ppm
|
|
Sterility assurance level (SAL) of 10-6 was achieved for Gulden eye spheres and conformers when processed according to STERRAD's instructions through a short full cycle in the STERRAD 100S Sterilization System.
|
|
|
|
|
|
|
|
 |